Masci, Paolo ; Zhang, Yi ; Jones, Paul ; Thimbleby, Harold ; Curzon, Paul

A Generic User Interface Architecture for Analyzing Use Hazards in Infusion Pump Software

pdf-format:
3.pdf (0.6 MB)


Abstract

This paper presents a generic infusion pump user interface (GIP-UI) architecture that intends to capture the common characteristics and functionalities of interactive software incorporated in broad classes of infusion pumps. It is designed to facilitate the identification of use hazards and their causes in infusion pump designs. This architecture constitutes our first effort at establishing a model-based risk analysis methodology that helps manufacturers identify and mitigate use hazards in their products at early stages of the development life-cycle.

The applicability of the GIP-UI architecture has been confirmed in a hazard analysis focusing on the number entry software of existing infusion pumps, in which the GIP-UI architecture is used to identify a substantial set of user interface design errors that may contribute to use hazards found in infusion pump incidents.

BibTeX - Entry

@InProceedings{masci_et_al:OASIcs:2014:4518,
  author =	{Paolo Masci and Yi Zhang and Paul Jones and Harold Thimbleby and Paul Curzon},
  title =	{{A Generic User Interface Architecture for Analyzing Use Hazards in Infusion Pump Software}},
  booktitle =	{5th Workshop on Medical Cyber-Physical Systems},
  pages =	{1--14},
  series =	{OpenAccess Series in Informatics (OASIcs)},
  ISBN =	{978-3-939897-66-8},
  ISSN =	{2190-6807},
  year =	{2014},
  volume =	{36},
  editor =	{Volker Turau and Marta Kwiatkowska and Rahul Mangharam and Christoph Weyer},
  publisher =	{Schloss Dagstuhl--Leibniz-Zentrum fuer Informatik},
  address =	{Dagstuhl, Germany},
  URL =		{http://drops.dagstuhl.de/opus/volltexte/2014/4518},
  URN =		{urn:nbn:de:0030-drops-45185},
  doi =		{10.4230/OASIcs.MCPS.2014.1},
  annote =	{Keywords: Infusion Pump, Hazard analysis, Use hazards, User Interface, Interactive software, Design errors}
}

Keywords: Infusion Pump, Hazard analysis, Use hazards, User Interface, Interactive software, Design errors
Collection: 5th Workshop on Medical Cyber-Physical Systems